Koehler Automated Flash Point Analyzers represent a perfect union of next-generation technology with traditional robust quality. The system software runs on an integrated processor PC running the latest Windows operating system. The 8.4” touch screen interface fully displays all operator parameters and results on a single screen. A three (3) position mechanical lift system for the cover and motor assembly is fully automated and software selectable: Open – Clean – Test for one touch positioning of the test cup. Comes standard with two flash detector systems including a thermocouple and ionization ring detection. Over 65,000 results of data can be stored on the local hard drive. Integrated Dual Fan System directly cools the test cup and the environment around test cup. Unlimited number of user programs including a quick test that safely tests from ambient, puts the flash point result into the EFP of the official run, and prompts user to refresh the sample, virtually assuring no fires ever occur. In addition, the instrument comes standard with an inert gas fire suppression system

Available Models

Koehler Automated Pensky-Martens Flash Point Tester
Compliance
ASTM D93 Procedure A, B and C; IP 34; ISO 2719; DIN EN 22719; JIS K2265; NF M 07-019
Detection System
Thermocouple and Ionization Ring
Stirring Rate
0 to 300 ± 5 RPM
Temperature Range
Ambient - 405°C
Power Requirements
110-240V 50/60Hz
Dimensions (cm)
27.63 x 44.45 x 66.04
Net Weight (kg)
27.22kg
Safety
Fire prevention, detection, and suppression systems standard
  • Conforms to ASTM D93 and related specifications
  • Flash Point Detection by Thermocouple and Ionization Ring
  • Electric or Gas ignition: Software selectable, User Friendly Manual Switching
  • Flash Point operation range between Ambient and 405°C
  • Integrated Dual Fan System directly cools test cup and surrounding
  • environment
  • 8.4” LCD Touch Screen Interface makes for easy viewing and navigation
  • Automatic Barometric Pressure Correction
  • Fire Suppression System floods instrument with inert gas in the event of a fire
Koehler Automated Cleveland Flash Point Tester
Compliance
ASTM D92; IP 36, 403; ISO 2592; AFNOR T60-118; EN 22592; DIN 51376
Detection System
Thermocouple and Ionization Ring
Power Requirements
115V 60Hz 1000W 230V 50Hz 1000W
Dimensions (cm)
53.5 x 26 x 50
Net Weight (kg)
24kg
Safety
Fire prevention, detection, and suppression systems standard
  • Conforms to ASTM D92 and related specifications
  • Simple automation routine for easy operation
  • Flash point operation between ambient and 400°C
  • Flash and Fire Point detection by Ionization Ring
  • Gas or electric ignition
  • Data Storage: 200 results
  • Automatic gas cut-off at the end of the test
  • Equipped with Emergency Safety Switch
  • Chemical
  • Education & Academics
  • Environmental
  • Oil & Gas
Koehler Instrument CompanyProviding quality testing instrumentation and technical support services for research and testing laboratories has been our specialty since 1925. Meeting your testing needs is the primary focus of our business, which is why Koehler Instrument Company is a leading producer and supplier of petroleum, synfuels and petrochemical instrumentation worldwide. At Koehler, we pride ourselves in innovation. In a time of continuous technological advancement and transition we are constantly implementing new and improved ways to surpass the needs of an ever-evolving industry. Koehler products are backed by our staff of technically knowledgeable, trained specialists who are experienced in both petroleum products testing and instrument service, which is carried out either on site or at a Koehler service center.