Technical Specifications

Particle size 0.5 µm - 1000 µm
Measurement Size, shape, transparency, count, location
Dispersion pressure 0.5 bar - 5 bar
Disp. pressure precision 0.1 bar
Injection time 5 ms - 100 ms
Sample settling time 1 s - 600 s
Available models
Malvern Morphologi_G3
Technical Specifications
Technology
Static automated imaging
Particle size
0.5 µm - 1000 µm
Measurement
Size, shape, transparency, count, location
Dispersion pressure
0.5 bar - 5 bar
Disp. pressure precision
0.1 bar
Injection time
5 ms - 100 ms
Sample settling time
1 s - 600 s
Microscope- Detector
5M pixel 2592 × 1944 colour CCD array
Microscope- Pixel width
2.78 µm × 2.78 µm
Microscope- Optical system
Nikon CFI 60 brightfield / darkfield system
Microscope- Lens
2.5× : 13 µm - 1000 µm5× : 6.5 µm - 420 µm10× : 3.5 µm - 210 µm20× : 1.75 µm - 100 µm50× : 0.5 µm - 40 µm
Weight and dimensions- Dimensions
440 mm × 760 mm × 700 mm
Weight and dimensions- Weight
80 kg
Operating environment- Power
100 V - 240 V, 50/60 Hz, 5.0 A
Operating temperature
10 to 40°C
Storage Operating environment- temperature
-20°C to 50°C
Operating environment- Humidity
-20°C to 50°C
Operating environment- Regulatory testing
CE, LVD, VCCI, C-Tick, FCC

Advanced Particle Characterization Made Easy

Malvern Morphologi G3 is an advanced yet easy-to-use particle characterization tool for the measurement of particle size and particle shape from 0.5 microns to several millimeters.

This single instrument offers the flexibility required for R&D and troubleshooting applications, as well as the user-independent results and validation required for automated QC analysis. The technique is often used in conjunction with laser diffraction particle sizing to gain a deeper understanding of product or process behavior.

Malvern Morphologi G3-ID
Technology
Static automated imaging with Raman Spectroscopy
Particle size
1 µm - 1000 µm
Measurement
Size, shape, transparency, count, location, chemical identification
Dispersion pressure
0.5 bar - 5 bar
Disp. pressure precision
0.1 bar
Injection time
5 ms - 100 ms
Sample settling time
1 s - 600 s
Detector
5M pixel 2592 × 1944 colour CCD array
Pixel width
2.78 µm × 2.78 µm
Optical system
Nikon CFI 60 brightfield / darkfield system
Standard particle size
2.5× : 13 µm - 1000 µm5× : 6.5 µm - 420 µm10× : 3.5 µm - 210 µm20× : 1.75 µm - 100 µm50× : 0.5 µm - 40 µm
Fully automated particle size
2.5× : 100 µm - 1000 µm5× : 25 µm - 420 µm10× : 20 µm - 210 µm20× : 20 µm - 100 µm50× : 1 µm - 40 µm
Spectral performance
150 cm⁻¹ to 1850 cm⁻¹, 6 cm⁻¹ resolution
Wavelength
785 nm
Power
<500 mW (spectrometer)>10 mW nominal (at sample)
Dimensions
440 mm × 760 mm × 700 mm (main instrument)584 mm × 448 mm × 203 mm (main instrument)
Weight
80 kg (main instrument)27.7 kg (spectrometer)
Operating environment- Power
100 V - 240 V, 50/60 Hz, 5.0 A
Operating temperature
15 to 30°C
Operating environment- Storage temperature
-15°C to 30°C
Operating environment- Humidity
0% - 95%

Automated Measurement of Particle Size, Shape and Chemical Identity

The Morphologi G3-ID offers a unique capability, combining the automated static imaging features of the Morphologi G3 with chemical identification of individual particles using Morphologically-Directed Raman Spectroscopy to enable the measurement of component-specific particle size and particle shape distributions.

  • Automated SOP control for unattended operation and reproducible results, boosting productivity

  • Excellent microscope images, accurate characterization

  • Versatile sample presentation accessories for measuring suspended and filtered samples

  • Advanced manual microscope control mode provides increased value as a shared laboratory resource

  • Rapid automatic particle counting on membrane filters

  • Powerful and intuitive software interface for easy visual and statistical interpretation of data

  • Aerospace
  • Agriculture
  • Automotive
  • Biopharmaceutical and Biotechnology
  • Chemical
  • Construction
  • Cosmetics & Personal Care
  • Defense
  • Die & Mould
  • Education & Academics
  • Electronics
  • Energy
  • Environmental
  • Food & Beverage
  • Luxury
  • Machinery & Manufacturing
  • Medical
  • Mining & Minerals
  • Oil & Gas
  • Optical
  • Paper & Wood
  • Pharmaceutical
  • Polymers, Plastics and Rubbers
  • Powders & Pigments / Coating
  • Railway
  • Semiconductor, Solar & Electronics
  • Service
  • Textile
  • Wholesale & Retail

Automated particle size, shape & chemical identity measurement with Morphologi G3-ID

The Morphologi G3-ID for automated measurement of particle size, particle shape and chemical identity.

http://bit.ly/VLyvMH - The Morphologi G3-ID provides a unique capability, combining automated static imaging features of the Morphologi G3 with chemical identification of individual particles using Raman spectroscopy to enable the measurement of component specific particle size and particle shape distributions. This fully automated instrument is designed to allow both particle characterization scientists with limited spectroscopy experience and more experienced spectroscopists to get an in-depth understanding of their particulate samples.

  • Cambodia
  • Indonesia
  • Laos
  • Malaysia
  • Myanmar
  • Philippines
  • Singapore
  • Thailand
  • Vietnam
Malvern Instruments LimitedMalvern systems are used to measure particle size, particle shape, zeta potential, protein charge, molecular weight, mass, size and conformation, microcalorimetry, rheological properties and for chemical identification, advancing the understanding of dispersed systems across many different industries and applications.
[Morphologi g3, Morphologi G3 SE, Morphologi g3-id, particle imaging, raman scattering, raman spectroscopy, ingredient specific, Morphologi g3se-id, Morphologi, Malvern]