Please leave a message
Designed specially for cost-effective laboratory thermostatting, LAUDA Alpha Cooling Thermostats strike a great balance between performance and value. A working temperature range between -25 and 100°C allows LAUDA Alpha Cooling Thermostats to cover most of basic laboratory thermostatic applications. LAUDA Alpha Cooling Thermostats' exceptional cost-efficiency is achieved by minimizing unnecessary features, with more focus and user-friendliness and reliability.
LAUDA Alpha Cooling Thermostats are suitable to be operated with non-flammable liquids such as water and water/glycol, and for both internal and external thermostatting applications. LAUDA Alpha Cooling Thermostats are also equipped with the capability of 1-point calibration, and this can be carried out easily by the user.
LAUDA Alpha Cooling Thermostats are designed with cooling technology which enables exceptional cooling performance through the entire temperature range from -25°C up to 100°C. A drain connection on the rear side of the devices empties the heat transfer liquid. Standard equipment includes bath covers and pump kits.
Working temperature range | -25-100°C |
---|---|
Temperature stability | 0.05 ±K |
Heater Power | 1.5 kW |
Pump pressure max. | 0.2 bar |
Pump flow max.* | 15 L/min |
Bath depth | 160 mm |
Technical Specifications
|
|
---|---|
Working temperature range
|
-25-100°C
|
Temperature stability
|
0.05 ±K
|
Heater Power
|
1.5 kW
|
Cooling output at 20°C
|
0.225 kW
|
Pump pressure max.
|
0.2 bar
|
Pump flow max.*
|
15 L/min
|
Bath volume
|
5-7.5 L
|
Bath Opening
|
165 mm × 177 mm
|
Bath Depth
|
160 mm
|
Technical Specifications
|
|
---|---|
Working temperature range
|
-25-100°C
|
Temperature stability
|
0.05 ±K
|
Heater Power
|
1.5 kW
|
Cooling output at 20°C
|
0.325 kW
|
Pump pressure max.
|
0.2 bar
|
Pump flow max.*
|
15 L/min
|
Bath volume
|
9.5-14.5 L
|
Bath Opening
|
300 mm × 203 mm
|
Bath Depth
|
160 mm
|
Technical Specifications
|
|
---|---|
Working temperature range
|
-25-100°C
|
Temperature stability
|
0.05 ±K
|
Heater Power
|
1.5 kW
|
Cooling output at 20°C
|
0.425 kW
|
Pump pressure max.
|
0.2 bar
|
Pump flow max.*
|
15 L/min
|
Bath volume
|
14-22 L
|
Bath Opening
|
350 mm × 277 mm
|
Bath Depth
|
160 mm
|
Easy and intuitive menu navigation and legible display values
Low level protection with acoustic and visual alarm to shut off heater and pump when heat transfer liquid is too low
Timer function to automatically shut down after preset time
Automatic compressor control supplies strong cooling output only when needed improves cost efficiency and equipment life
Fitted with screw clamp to allow easy change to different bath vessels with wall thickness of up to 30 mm
Variable flow rates via inserts and fittings adapt rate of circulation to the size of the bath