The KT 200 Kjeltec™ Distillation unit provides a simple and reliable solution for safe and semi-automatic distillation for protein determination. The possibility to pre-program alkali addition and distillation time supports the production of accurate results independent of operator.
The highly versatile Kjeldahl distillation unit uses the global reference method for automated/semi-automated protein analysis with the highest safety standards for use in any laboratory.
KT 200 Kjeltec™ | Kjeltec™ 8100 | Kjeltec™ 8200 | Kjeltec™ 8400 | |
---|---|---|---|---|
Value | Simple, safe and reliable analysis according to official standards for accurate results independent of operator. | Free-up lab resources and improve test throughput with semi-automated distillation that is safe, simple and reliable and according to official standards. | Future proof Kjeldahl. Take advantage of an versatile automated distillation system with the option to upgrade to a fully automated distillation and titration system according to your laboratory demands. | Fully automated (distillation, colorimetric titration, calculation). Upgradeable to auto sampler. Easy data handling with Compass software. |
Sample type | Food, feed, feed ingredients, forage, pet food, grain, cereal, oil seeds, meat, meat products, milk, dairy products, beer and beer ingredients. | Food, feed, feed ingredients, forage, pet food, grain, cereal, oil seeds, meat, meat products, milk, dairy products, beer and beer ingredients. | Food, feed, feed ingredients, forage, pet food, grain, cereal, oil seeds, meat, meat products, milk, dairy products, beer and beer ingredients. | Feed, feed ingredients, forage, pet food, grain, cereal, oil seeds, meat, meat products, milk, dairy products, beer and beer ingredients. |
Parameters | Nitrogen, protein, Non Protein Nitrogen (NPN), and volatiles (SO2, TVBN and more). | Nitrogen, protein, Non Protein Nitrogen (NPN) and volatiles (SO2, TVBN and more). | Nitrogen, protein, Non Protein Nitrogen (NPN) and volatiles (SO2, TVBN and more). | Nitrogen, protein, Non Protein Nitrogen (NPN) and volatiles (TVBN and more). |
Technical specifications
|
|
---|---|
Performance data at 230V
|
|
Analysis time
|
4 minutes at 30 mg N
(7 minutes at 200 mg N)
|
Distillation capacity
|
~ 30 ml/min
|
Measuring range
|
0.1 – 200 mg N
|
Reproducibility
|
1% RSD (including the digestion step)
|
Recovery
|
> 99.5% at nitrogen levels between 1 – 200 mg N
|
Installation Requirements
|
|
Power supply
|
200 – 240 V 50 – 60 Hz
|
Power consumption
|
2 200 W
|
Water consumption
|
2 l/min at water temp of 30°C (During distillation only)
1 l/min at water temp of 15°C
|
Weight net
|
25 kg
|
Dimensions (W x D x H)
|
300 × 280 × 770 mm
|
A high accuracy bellows pump gives long term stability in delivery of the alkali. The unit is equipped with a wide range of safety features including safety door, plus level pins and overpressure sensors for the steam generator. The door completely covers the area in front of the operator and the unit cannot be operated with the door open. The drip tray can easily be removed for cleaning.
Official and accurate procedures (ISO, AOAC, EPA, DIN) simplify validation.
Technical specifications
|
|
---|---|
Performance data
|
|
Analysis time
|
3.5 minutes at 30 mg N (6.5 minutes at 200 mg N)
|
Distillation capacity
|
~ 40 ml/min
|
Measurement range
|
0.1 – 200 mg N
|
Reproducibility
|
1% RSD (including the digestion step)
|
Recovery
|
> 99.5% at nitrogen levels between 1 – 200 mg N
|
Installation requirements
|
|
Power supply
|
200 – 240 V 50 – 60 Hz
|
Tolerable voltage variation
|
± 10 %
|
Power consumption
|
2 200 W
|
Water consumption
|
(during distillation only) 3 l/min at water temp of 25°C *1 l/min at water temp of 15°C
* Circulating Cooler recommended at >25°C
|
Weight
|
35 kg
|
Dimensions (W x D x H)
|
480 × 580 × 690 mm
|
A fast tube draining system removes the need for handling hot reagents after distillation. Interlocked safety door which must be closed with a tube in place before the unit will operate. Cooling water flow is controlled automatically and a temperature sensor stops the unit if the distillate is too hot. Sensors monitor tube in place, safety doors, reagent levels, water level and over-pressure in the steam generator providing safe use.
Technical specifications
|
|
---|---|
Performance data at 230 V
|
|
Analysis time
|
3.5 minutes at 30 mg N (6.5 minutes at 200 mg N)
|
Distillation capacity
|
~ 40 ml/min
|
Measuring range
|
0.1 – 200 mg N
|
Reproducibility
|
1% RSD (including the digestion step)
|
Recovery
|
> 99.5% at nitrogen levels between 1 – 200 mg N
|
Installation requirements
|
|
Power supply
|
200 – 240 V 50 – 60 Hz
|
Tolerable voltage variation
|
± 10 %
|
Power consumption
|
2 200 W
|
Water consumption
|
(during distillation only) 3 l/min at water temp of 25°C
*1 l/min at water temp of 15°C
*Circulating Cooler recommended at >25°C
|
Weight net
|
35 kg
|
Dimensions: (W×D×H)
|
480 × 580 × 690 mm
|
A fast tube draining system removes the need for handling hot reagents after distillation. Interlocked safety door which must be closed with a tube in place before the unit will operate. Cooling water flow is controlled automatically and a temperature sensor stops the unit if the distillate is too hot. Sensors monitor tube in place, safety doors, reagent levels, water level and over-pressure in the steam generator providing safe use.
Technical specifications
|
|
---|---|
Performance data at 230 V
|
|
Autosampler capacity
|
Kjeltec™ sampler 8420, 1 rack with 20 Kjeldahl tubes 250 ml or 400 ml
Kjeltec™ sampler 8460, 3 racks with 20 Kjeldahl tubes 250 ml or 400 ml
|
Analysis time
|
3.5 minutes at 30 mg N (6.5 minutes at 200 mg N)
|
Distillation capacity
|
~ 40 ml/min
|
Measuring range
|
0.1 – 200 mg N
|
Reproducibility
|
1% RSD (including the digestion step)
|
Installation requirements
|
|
Power supply
|
200 – 240 V 50 – 60 Hz
|
Tolerable voltage variation
|
± 10 %
|
Power consumption
|
2 200 W
|
Water consumption
|
(during distillation only) 3 l/min at water temp of 25°C
*1 l/min at water temp of 15°C
* Circulating Cooler recommended at >25°C
|
Weight net
|
35kg
|
Dimensions: (W×D×H)
|
480 × 580 × 690 mm
|
The superior performance of a KjeltecTM titration system is based on colorimetric measurement of a precise chemical reaction for end-point detection. Accuracy is assured time after time with the additional benefits of long-term stability, wide working range, speed and simplicity without the need for repeated and costly calibration procedures.
A fast tube draining system removes the need for handling hot reagents after distillation. Interlocked safety door which must be closed with a tube in place before the unit will operate. Cooling water flow is controlled automatically and a temperature sensor stops the unit if the distillate is too hot. Sensors monitor tube in place, safety doors, reagent levels, water level and over-pressure in the steam generator providing safe use.
An optional 20 or 60 place Autosampler provides the benefits of automation even at lower sample throughputs. Just load your sample racks directly from the digestion block and KjeltecTM will perform accurate analysis unattended for more than four hours.
The patented SAfE* technology improves safety by a mixing procedure, reducing the exothermic reaction between alkali and acid during distillation and removing the need for manual pre-dilution.
*Steam Addition for Equilibration
KT 200 Kjeltec™ Distillation Unit
Kjeltec™ 8100 Manual Distillation Unit
Kjeltec™ 8200 Auto Distillation Unit
Kjeltec™ 8400 fully automated Kjeldahl analysis
FOSS creates end-to-end solutions that secure and improve food quality. From raw material to finished product. Our analysis instruments refine measurements into information management that enables businesses to run intelligent data-driven productions with less waste and bigger yields. Controlling cost and quality across all sectors and value chains, we help food and agricultural producers limit the number of human errors, scale their business faster and reduce manual labor as well as labor costs. We call it analytics beyond measure.
We cover these food and agricultural industries: Dairy, Feed and Forage, Grain, Flour Milling & Oilseed Processing, Laboratories, Meat, Raw Milk Testing and Wine.
Over the last 60 years, we have made it our business to define the standard of analytical solutions for the food and agricultural industry. We pioneered advanced analysis for the challenging production environments in the 1960’s, and we were the first to integrate chemical composition analysis in process lines. Now our resolve and ambition have also made us the international benchmark in the field of chemometrics and real time analytics.