This Toroidal Conductivity sensor head is part of the AdvantEDGE probe system that allows a direct connection to a PLC for seamless integration with industrial control systems. Use any computer to display data, calibrate and customize the measurement without an intermediate analyzer electronics box. A LD500 local display can be used if desired. Sensor heads are pre-calibrated and can be replaced or exchanged with any other type of sensor without taking the system down. Save space, time and money.
Measurement System Performance | Range: 0 to 2,000,000 uS/cm (2 S/cm)Resolution: 4.5 significant digitsAccuracy: 0.1% of readingStep Response Time: 90% in 30 secondsNote: Typical at 25°CPerformance unaffected by cable length |
---|---|
Operational Environment | Temperature Range: -5°C to 95°CMaximum Pressure: 85 psi @ 50°CMaximum Flow Rate: 10 ft/second |
Power Requirements | Voltage Range: 10 to 30 VDCMaximum Power: 200 mWTypical Power: 120 mWNote: Class II DC power supply required |
Approvals and Ratings | Immunity & Emissions: CE (EN61326 and EN61326:1998)Safety: CE (EN 61010-1) UL 508Hazardous Locations: UL Class I, Division 2 |
Construction | Cell Constants: 0.1 for Resistivity 1.0 and 5.0 for ConductivityO-Rings: Viton® (other materials available)Electrode Material: Titanium (Standard) or 316 Stainless SteelWeight: 1.2 lbs |
Units of Measure | Measurement Units: μS/cm, TDS, %, MΩ • cTemperature Units: °C, °F |
Calibration | Zero: In dry airSpan: 1 pointTemperature: 1 and 2 pointNote: Conductivity and temperature are precalibrated atthe factory |
Other Configuration Options | Sensor Filter: 0 to 100 secondsTemperature Filter: 0 to 100 secondsConcentration Table: User 10-point |