Designed to provide the accuracy demanded by today’s scientist, the Micro100 Laboratory Turbidimeter is ready to meet the challenge. Features like the Auto Alert Calibration Prompt and the Quick Connection Lamp Module, allow for simple calibration and low maintenance. With a resolution of 0.01 NTU at low turbidity readings and an extended range to 1,000 NTU, the Micro100 is an ideal and affordable tool for research as well as routine analytical measurements.
Technical Specifications
|
|
---|---|
Principal of operation
|
Nephelometric non-ratio
|
Range
|
Auto ranging from 0 - 1000 NTU
|
Accuracy
|
± 2% of reading plus 0.01 NTU
|
Resolution
|
0.01 NTU (0 - 9.99 NTU)0.1 NTU (10 - 99.9 NTU)1 NTU (100 - 1000 NTU)
|
Response Time
|
Less than 6 seconds
|
Sample size
|
30 ml (27 ml minimum)
|
Light source
|
Quick connect tungsten filament or infrared
|
Temperature
|
32° - 122°F (0° - 50°C)
|
Outputs
|
RS-232 serial port
|
Dimensions
|
10.75 x 10 x 3.75 inches (27.3 x 25.4 x 9.5 cm)
|
Weight
|
2.9 lbs. (1.32 kg)
|
Safety Rating
|
UL, CSA, CE
|
Weight
|
7.0 lbs. (3.18 kg)
|
Certification
|
UL, CSA, CE, USEPA, ISO7027
|
Micro100 IR Laboratory Turbidimeter, 0 - 1000 NTU, Universal Power 100-240V 50-60 Hz Includes a complete set of calibration standards (0.02, 10 and 1000 NTU), two measuring cuvettes with light shield cap, operator's manual and a universal power adapter that operates at any voltage between 100-240VAC.
Technical Specifications
|
|
---|---|
Principal of operation
|
Nephelometric
|
Range
|
0 - 10,000 NTU0 - 2,450 EBC0 - 10,000 FNU0 - 10,000 FAU
|
Ratio (color correction)
|
Full time ON or OFF
|
Accuracy
|
± 2% of reading plus 0.01 NTU (0 to 1000 NTU)± 5% of reading (1000 to 4000 NTU)± 10% of reading (4000 to 10,000 NTU)
|
Resolution
|
0.0001 NTU on lowest range
|
Response time
|
Less than 6 seconds
|
Sample size
|
30 ml
|
Light source
|
Quick connect tungsten filament or infrared
|
Operating temperature
|
32° - 122°F (0° - 50°C)
|
Air purge
|
Connection for external dry air supply
|
Outputs
|
RS-232 serial port (bidirectional)
|
Shipping dimensions
|
14 x 12 x 12 inches (35.5 x 30.4 x 30.5 cm)
|
Micro100 IR Laboratory Turbidimeter, 0 - 1000 NTU, Universal Power 100-240V 50-60 Hz Includes a complete set of calibration standards (0.02, 10 and 1000 NTU), two measuring cuvettes with light shield cap, operator's manual and a universal power adapter that operates at any voltage between 100-240VAC.