Need support?

Please leave a message

×
Seta - Pensky-Martens Flash Point Tester - PM-93

The Seta PM-93 is a fully automated Pensky-Martens Closed Cup Flash Point Tester which combines strict method conformance with the latest control technology and safety systems to provide the next generation of flash point instruments.

An intuitive graphic user interface and 8.4” SVGA colour touch screen operation provide unparalleled ease of use with real time display of test progress. Test profiles, operator names, test methods and storage of unlimited results are provided through the embedded operating system. Test results can be emailed, saved to a memory stick, output to LIMS or produced in hard copy with the printer option.

Technical specifications

Ignition system: Electric or gas
Flash detection: Thermal
Cooling: Forced air (integral fan)
Heating Rate: Fast heating rate (>10°C/min) and standard; 5.5°C/min; 3°C/min; 1.3°C/min; 1°C/min
Temperature range: Ambient +5°C to 400°C
Barometric pressure correction: Automatic correction with built-in pressure sensor
Internal Memory: 30 programmable test profiles, operator names, test methods and unlimited tests results
Fire Detection: Thermal with optional integral Fire Extinguisher
Dimensions (H xW x D) / weight: 38.5 x 24 x 50 cm / 25 kg
Voltage 230 Vac +/-15% or 110 Vac +/-15%. Frequency 50/60 Hz. Auto-sensing - no selector switch required.

Key features

  • New design, latest enhanced operation
  • Fully conforms to test methods with 4 heating rates available
  • Simple ‘3-step to test’ operation
  • Unique patented ‘SafeFlash’ flame extinguisher system
  • Large 8.4” Resistive touch screen menu
  • Single action lifting arm operation
  • Pre-dip safety mode
  • Programmable with 30 test profiles, test methods & sample identities
  • Test memory = 2000 results
  • Gas or electronic ignition using proven and long lasting Seta ‘Ignite’ technology
  • Compact footprint

Testing methods and details

BS EN ISO 2719, BS 2000-34 Methods of test for petroleum and its products. Flash point by Pensky-Martens closed tester
ASTM D93 Standard test methods for flash-point by Pensky-Martens closed cup tester
IP 34 Determination of flash point - Pensky - Martens closed cup method (ISO 2719:2002)

Pensky-Martens Flash Point Testing by PM-93 Pensky-Martens Flash Point Tester for ASTM D93 | IP 34 | ISO 2719 by Seta PM-93

  • Environmental
  • Oil & Gas
  • Pharmaceutical
Stanhope-SetaStanhope-Seta design and manufacture quality control instruments used to measure the physical characteristics that determine product quality and consistency. The 'Seta' range includes test equipment for the hydrocarbons industry as well as many other laboratory test applications for quality control, calibration and analysis.
[]